Microchannel plate (MCP) photomultiplier tubes (PMTs) are frequently used in experimental diagnostics, where they are operated in single-pulse current measurement mode. However, considering the significant amplitude fluctuations in the measured signal, the resulting output signal from the MCP-PMT is inevitably distorted by gain saturation. Therefore, understanding the correlation between the MCP-PMT output signal and gain saturation is critical in assessing the extent of output signal distortion and determining the MCP-PMT saturation level. This knowledge allows for a more precise assessment of the input signal's features. In this paper, we present an experimental method for restoring the initial waveform from the saturated MCP-PMT signal. To correct the amplitude-drop caused by gain saturation, our technique involves calibrating the MCP-PMT's relative gain as a function of the accumulated output charge using a square-wave light source. We then applied this approach to restore a similar to 500 ns saturated pulse from a double-layer 10 mm diameter MCP-PMT. The restored signal showed a deviation of less than 6% from the reference waveform, which validates the effectiveness of the technique.