Radiation doses in extraoral bitewing radiography compared with intraoral bitewing and panoramic radiography

ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY(2024)

引用 0|浏览2
暂无评分
摘要
Objectives. We compared the effective dose (E) and thyroid equivalent dose of 2 extraoral bitewing (EOBW) units and compared Study Design. Child and adult anthropomorphic phantoms with dosimeters were used to evaluate Orthophos SL, Rayscan a+, and 1 intraoral unit using rectangular and circular collimation. Extraoral bitewing thyroid equivalent dose was assessed without and with thyroid shielding. Results. Child and adult E values of EOBW were lower with Orthophos (3.6 and 8.6 mu Sv) than with Rayscan (28.1 and 30.2 mu Sv). For IOBW, E was lower with rectangular vs circular collimation for child (7.0 vs 11.8 mu Sv) and adult (4.6 vs 14.2 mu Sv). E values of EOBW were lower than PAN for Orthophos. The IOBW E was lower than Rayscan EOBW for child (<= 11.8 vs 28.1 mu Sv) and adult (<= 14.2 vs 30.2 mu Sv). Adult E for rectangular IOBW (4.6 mu Sv) was lower than EOBW with Orthophos (8.6 mu Sv) and Rayscan (30.2 mu Sv). Thyroid shielding reduced EOBW thyroid equivalent dose with Rayscan in the adult from 190.7 to 89.0 mu Sv. Conclusion. Orthophos provides significantly lower EOBW E than Rayscan, thus EOBW recommendations must be unit specific. For children, Orthophos EOBW could be an alternative to IOBW, for which rectangular collimation is recommended. Thyroid shielding reduced adult Rayscan equivalent dose but added imaging artifacts. (Oral Surg Oral Med Oral Pathol Oral Radiol 2024;137:182-189)