Sputter sample preparation for ion beam delivery of radium-223 at ATLAS.

J McLain, M Gott, J Greene, R Scott,R Vondrasek

The Review of scientific instruments(2023)

引用 0|浏览2
暂无评分
摘要
A radium-223 ion beam was delivered to an experiment from the electron cyclotron resonance ion source, ECR2, at the Argonne Tandem Linac Accelerator System (ATLAS). The radium-223 material was in a nitrate salt form within a vial, prior to being converted to a usable sputter sample. The sputter sample was produced using a new sample preparation method, where the radium nitrate was dissolved into a solution and pipetted onto pressed aluminum powder. This sample was then allowed to dry, distributing the radium-223 material throughout the sputter sample. Ion source operation using the radium sputter sample is described with the operating parameters listed. The intensity and energy requirements for this ion beam were 1 × 106 particles/s and 1.07 GeV, respectively. Because the intensity is relatively low compared to most experiments at ATLAS, previously developed accelerator mass spectrometry methods were used Scott et al. [Rev. Sci. Instrum. 87, 02A732 (2016)] to avoid the need for tuning of the low-intensity beam of interest. Handling of the radium material, as well as loading and unloading of the sputter sample from ECR2, required collaboration with Health Physics. Procedures were used and dry runs were carried out before, during, and after the experiment to ensure the safety of the workers. The processes used and lessons learned are described within.