Gas chromatography-flame ionization detector for sweat based COVID-19 screening.

Analytica chimica acta(2023)

引用 0|浏览2
暂无评分
摘要
Simple approach for rapid screening of corona virus disease 2019 (COVID-19) has been developed. This applied gas chromatography-flame ionization detector (GC-FID) analyzing the potential compound marker in sweat samples obtained from COVID-19 positive and negative volunteers in Bangkok, Thailand. The samples were collected by using cotton rods for 15 min, heated at 90 °C for 5 min, and the volatile compounds in the headspace (HS) were injected (5.00 mL) at 150 °C and separated within 13.7 min. The marker peak was tentatively identified as p-cymene by the authentic standard injection and comparison with the GC-mass spectrometry (GC-MS) and comprehensive two-dimensional GC (GC × GC)-MS analysis. Possible mechanisms for the presence of p-cymene were proposed. The marker peak area thresholds were then varied and optimized via construction of the receiver operating characteristic (ROC) curve. With the optimum threshold, the established method offered the accuracy, sensitivity and specificity of 96 %. This method was insignificantly affected (p-value >0.05) by genders, body mass indices, ages, and use of deodorants as well as the p-cymene containing food. However, the performance could be affected by the population with personal hygiene or experiencing the microbiomes producing p-cymene.