Vancomycin and Gentamicin Removal with the HA380 Cartridge during Experimental Hemoadsorption

Blood purification(2023)

引用 0|浏览6
暂无评分
摘要
Introduction: Hemoadsorption has emerged as an adjunctive therapy for sepsis, but its impact on antibiotic levels remains poorly defined. We conducted an in vivo experimental study to investigate the removal of vancomycin and gentamicin during hemoadsorption using the HA380 cartridge, a novel styrene-divinylbenzene copolymer cartridge. Methods: Six surgically prepared sheep were administered 2 g of vancomycin and 400 mg of gentamicin over 30 min, followed by a continuous infusion of vancomycin (20 mg/h). Hemoadsorption was implemented with a styrene-divinylbenzene copolymer HA380 cartridge at a blood flow of 120 mL/min. The removal ratio, sorbent-based clearance, and the mass removal rate were calculated for each time point. Results: The mean 10-min vancomycin removal ratio exceeded 90% and declined to 68.0% at 30 min; 52.8% at 60 min, and 28.0% by 4 h. Due to constant plasma flow, clearance varied proportionally with the removal ratio. Over 4 hours, the total mass removal was 556 mg (SD 106.3). For gentamicin, the mean 10-min removal ratio was 96.9% and the final ratio at 4 h remained 53.0%, with clearances changing proportionately. The total mass removal of gentamicin was 138 mg (SD 26.6) over 4 h. The sorbent-based clearance of vancomycin was significantly lower than that of gentamicin (P-group < 0.0001). Conclusion: The novel HA380 sorbent cartridge appears safe and achieves significant vancomycin and gentamicin removal over a four-hour period. This information can be used by clinicians to guide their prescription and consider the additional dosing of at least an extra 25-35% amount in patients receiving HA380 hemoadsorption therapy during sepsis.
更多
查看译文
关键词
gentamicin removal,vancomycin,ha380 cartridge
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要