Particular issues to be considered in small field dosimetry for TrueBeam STx commissioning

APPLIED RADIATION AND ISOTOPES(2023)

Cited 0|Views5
No score
Abstract
This study aims to report the relevant issues concerning small fields in the commissioning of a TrueBeam STx for photon energies of 6MV, 10MV, 6FFF, and 10FFF. Percent depth doses, profiles, and field output factors were measured according to the beam model configuration of the treatment planning system. Multiple detectors were used based on the IAEA TRS-483 protocol as well as EBT3 radiochromic film. Analytical Anisotropic and Acuros XB algorithms, were configured and validated through basic dosimetry comparisons and end-to-end clinical tests.
More
Translated text
Key words
Small field dosimetry,TRS-483,Field output factor,TrueBeam STx,Acuros algorithm