Sub-aperture ultrafast volumetric ultrasound imaging for fully sampled dual-mode matrix array

ULTRASONICS(2024)

引用 0|浏览14
暂无评分
摘要
Fully sampled dual-mode matrix array ultrasound transducer is capable of performing imaging and therapeutic ultrasound in three dimensions (3D). It is a promising tool for many clinical applications because of its precise multi-focus therapy with imaging guidance by itself. Our team previously designed a 256-element fully sampled dual-mode matrix array transducer, while its imaging quality needs to be further improved. In this work, we propose a high-contrast sub-aperture volumetric imaging strategy to improve the imaging quality of the dual mode matrix array. We first analyzed the effect of various parameters of sub-aperture imaging on the imaging quality by Field II. Based on the optimized parameters, we compared the resolution and signal to noise ratio (SNR) of sub-aperture imaging with those of full aperture imaging on phantoms and rabbit brain. The experimental results showed the proposed sub-aperture imaging method could obtain a comparable resolution to full aperture imaging. Moreover, the average intensity of noise signal near the wire phantom decreased by about 5 dB and the SNR of tissue phantom image increased by 8 %. The proposed sub-aperture imaging method also enabled clearer and more accurate imaging of the rabbit brain. The obtained results indicate the proposed sub-aperture imaging is a promising method for practical use of a fully sampled dual-mode matrix array for volumetric ultrasound imaging.
更多
查看译文
关键词
Volumetric ultrasound imaging,Sub-aperture imaging,Dual-mode matrix array,Imaging guidance