Agreement of Computerized QT and QTc Interval Measurements Between Both Bedside and Expert Nurses Using Electronic Calipers.

The Journal of cardiovascular nursing(2023)

Cited 0|Views15
No score
Abstract
Computerized QT/QTc measurements calculated by bedside monitor software and measurements performed by nurses were in close agreement; statistically significant differences were found, but differences were less than 20 milliseconds (on-half of a small box), indicating no clinical significance. Computerized measurements may be a suitable alternative to nurse-measured QT/QTc. This could reduce inaccuracies and nurse burden while increasing adherence to practice recommendations. Further research comparing computerized QT/QTc from bedside monitoring to standard 12-lead electrocardiogram in a larger sample, including non-ICU patients, is needed.