Validation of the Iowa Head-Mounted Open-Source Perimeter

Translational vision science & technology(2023)

引用 0|浏览2
暂无评分
摘要
Purpose: To assess the validity of visual field (VF) results from the Iowa Head-Mounted Display (HMD) Open-Source Perimeter and to test the hypothesis that VF defects and test-retest repeatability are similar between the HMD and Octopus 900 perimeters. Methods: We tested 20 healthy and nine glaucoma patients on the HMD and Octopus 900 perimeters using the Open Perimetry Interface platform with size V stimuli, a custom grid spanning the central 26 degrees of the VF, and a ZEST thresholding algorithm. Historical data from the Humphrey Field Analyzer (HFA) were also analyzed. Repeatability was analyzed with the repeatability coefficient (RC), and VF defect detection was determined through side-by-side comparisons. Results: The pointwise RCs were 2.6 dB and 3.4 dB for the HMD and Octopus 900 perimeters in ocular healthy subjects, respectively. Likewise, the RCs were 4.2 dB and 3.5 dB, respectively, in glaucomatous patients. Limits of agreement between the HMD and Octopus 900 perimeters were +/- 4.6 dB (mean difference, 0.4 dB) for healthy patients and +/- 8.9 dB (mean difference, 0.1 dB) for glaucomatous patients. Retrospective analysis showed that pointwise RCs on the HFA2 perimeter were between 3.4 and 3.7 dB for healthy patients and between 3.9 and 4.7 dB for glaucoma patients. VF defects were similar between the HMD and Octopus 900 for glaucoma subjects. Conclusions: The Iowa Virtual Reality HMD Open-Source Perimeter is as repeatable as the Octopus 900 perimeter and is a more portable and less expensive alternative than traditional perimeters. Translational Relevance: This study demonstrates the validity of the visual field results from the Iowa HMD Open-Source Perimeter which may help expand perimetry access.
更多
查看译文
关键词
virtual reality,glaucoma,perimetry,visual field
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要