Removing biases in dual frequency comb spectroscopy due to digitizer nonlinearity

OPTICS EXPRESS(2023)

引用 0|浏览7
暂无评分
摘要
Operation of any dual-comb spectrometer requires digitization of the interference signal before further processing. Nonlinearities in the analog-to-digital conversion can alter the apparent gas concentration by multiple percent, limiting both precision and accuracy of this technique. This work describes both the measurement of digitizer nonlinearity and the development of a model that quantitatively describes observed concentration bias over a range of conditions. We present hardware methods to suppress digitizer-induced bias of concentration retrievals below 0.1%.