Determination of Metals by Portable Millisecond Pulsed Solution Cathode Glow Discharge (SCGD) Atomic Emission Spectroscopy (AES)

Peichao Zheng, Ningshen Wang,Jinmei Wang, Hafiz Imran Ahmad Qazi,Hongwu Tian,Daming Dong, Yuxin He,Meini Wu

ANALYTICAL LETTERS(2024)

引用 0|浏览4
暂无评分
摘要
In order to determine metal ions in aqueous solution, a millisecond pulsed solution cathode glow discharge (SCGD) source coupled with a portable spectrometer was developed for atomic emission spectrometry (AES). The millisecond pulsed SCGD source was operated with two high-voltage DC power supplies and a high-voltage pulse switch, specifically boosting the analyte excitation and sustain the glow discharge. The method to excite the plasma discharge contrasts with traditional approaches that employ a single power source. The influence of operating parameters, including pulsed voltage, pulsed frequency, duty ratio, and sample flow rate, were optimized to construct a system having good stability and improved excitation of metal ions. The optimum detection limits for Na, K, Li, Mg, Ag, Cs, Rb, and Sr were 0.17, 0.32, 0.38, 0.50, 2.65, 1.30, 0.22, and 74.02 & mu;g & BULL;L-1, respectively. Furthermore, mountain spring, Yangtze River, and reservoir water were spiked with Ag, Rb, and Cs; the recoveries ranged from 105% to 106%, 106% to 107%, and 107% to 108% indicating that the developed system has favorable analytical performance.
更多
查看译文
关键词
Atomic emission spectrometry (AES),elemental analysis,millisecond pulsed solution cathode glow discharge (SCGD),solution spectroscopy
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要