Analysis of Contact Allergens in Polyvinyl Chloride Examination Gloves in the United States

DERMATITIS(2024)

引用 0|浏览4
暂无评分
摘要
Background: Polyvinyl chloride (PVC) gloves are recommended as a safe alternative for patients with rubber accelerator allergy. However, allergic contact dermatitis to other chemicals in PVC gloves has been reported.Objective: To analyze single-use PVC medical examination gloves in the United States for the presence of potential contact allergens. Methods: Using liquid chromatography-mass spectrometry, 20 unique PVC gloves were analyzed in triplicate for 6 chemicals: benzisothiazolinone, bisphenol A, mono(2-ethylhexyl) maleate, tricresyl phosphate, triphenyl phosphate, and triphenyl phosphite. Results: All 20 PVC gloves contained detectable quantities of benzisothiazolinone (range, 0.001-1.48 parts per million [ppm]), bisphenol A (0.01-0.11 ppm), triphenyl phosphate (0.01-2.11 ppm), and triphenyl phosphite (0.001-0.22 ppm). Eighteen (90%) gloves contained mono(2-ethylhexyl) maleate (0.001-0.14 ppm) and 3 (15%) contained tricresyl phosphate (0.001-0.002 ppm). Conclusions: Known allergens were present in all 20 PVC gloves. However, the detected levels were mostly low and their relationship with sensitization and elicitation thresholds requires further study.