Symmetric-Reciprocal-Match Method for Vector Network Analyzer Calibration

IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT(2024)

引用 0|浏览3
暂无评分
摘要
This article proposes a new approach, the symmetric-reciprocal-match (SRM) method, for calibrating vector network analyzers (VNAs). The method involves using multiple symmetric one-port loads, a two-port reciprocal device, and a matched load. The load standards consist of two-port symmetric one-port devices, and at least three unique loads are used. However, the specific impedances of the loads are not specified. The reciprocal device can be any transmissive device, although a nonreciprocal device can also be used if only the one-port error boxes are of interest. The matched load is fully defined and can be asymmetric. We demonstrated the accuracy of the proposed method with measurements of coaxial standards using a commercial METAS traceable short-open-load-reciprocal (SOLR) calibration kit with verification standards. In addition, we presented a numerical Monte Carlo (MC) analysis considering various uncertainty factors. An advantage of the proposed method is that only the match standard is defined, whereas the remaining standards are partially defined, either through symmetry or reciprocity.
更多
查看译文
关键词
Calibration,microwave measurement,vector network analyzer (VNA)
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要