Influence of resident involvement on fluoroscopy time and ionizing radiation exposure in fluoroscopy-guided spinal procedures

PM&R(2024)

引用 0|浏览2
暂无评分
摘要
Background: Fluoroscopic guidance has become the standard for a variety of medical procedures. Mastering these techniques requires practice, which may entail additional radiation for patients and providers. Despite their widespread use, the literature examining factors influencing radiation exposure in fluoroscopically guided pain procedures is scarce.Objective: To evaluate the influence of resident involvement on radiation exposure during fluoroscopy-guided spinal interventions.Design: Single-center, observational study.Setting: Outpatient physiatry clinic in a teaching hospital.Patients: All patients who received cervical or lumbar facet block(s) (FBs), transforaminal epidural steroid injection(s) (TFESIs) without digital subtraction, or a caudal epidural (CE) during the study period were included.Interventions: Resident involvement in the procedures: absent, observing, or participating.Main Outcome Measures: Machine-indicated fluoroscopy time (seconds) and radiation dose (milligrays [mGy]).Results: Two hundred ninety six procedures were included: 188 FBs (58 cervical, 130 lumbar), 48 CEs, and 60 TFESIs. For lumbar FBs, fluoroscopy time and radiation dose increased significantly when residents performed them (meantime = 24.5 s, confidence interval [CI] = 20.4-28.7; meandose = 3.53 mGy, CI = 2.57-4.49) compared to when they observed (meantime = 9.9 s, CI = 8.1-11.7; meandose = 1.28 mGy, CI = 0.98-1.59) (mean difference: time = 14.63 s, CI = 9.31-19.94; dose = 2.25 mGy, CI = 1.17-3.33) and were absent during the procedure (meantime = 12.9 s, CI = 11.1-14.6; meandose = 1.65 mGy, CI = 1.40-1.89) (mean difference: time = 11.67 s, CI = 7.35-15.98; dose = 1.88 mGy, CI = 1.01-2.76). In the case of TFESIs, time, but not dose, increased significantly when residents observed (meantime = 39.1 s, CI = 30.7-47.6; meandose = 6.73 mGy, CI = 3.39-10.07) compared to when they were absent (meantime = 27.1 s, CI = 22.4-31.8; meandose = 4.41 mGy, CI = 3.06-5.76 (mean difference: time = 11.99 s, CI = 1.37-22.61; dose = 2.32 mGy, CI = -1.20-5.84). Finally, resident involvement did not significantly affect the outcomes for CEs (ptime = .032, pdose = .74) and cervical FBs (ptime = .64, pdose = .68).Conclusion: Resident participation affected lumbar FBs the most, with an increase in both fluoroscopy time and radiation dose.
更多
查看译文
关键词
fluoroscopy time,radiation exposure,resident involvement
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要