Performance of an active transcutaneous bone conduction implant, the Osia system, in high frequencies and hearing in noise in users converted from a Baha attract system: a prospective study

European archives of oto-rhino-laryngology : official journal of the European Federation of Oto-Rhino-Laryngological Societies (EUFOS) : affiliated with the German Society for Oto-Rhino-Laryngology - Head and Neck Surgery(2023)

引用 0|浏览5
暂无评分
摘要
Purpose Hearing rehabilitation through middle ear surgery can be challenging, particularly for patients with chronic otitis media, but new hearing devices offer opportunity to reach this goal. The aim was to compare hearing outcomes and quality of life in patients who were converted from a Baha Attract system to the Osia system. Methods Prospective observational study, in which each subject acted as their own control. Six patients who were converted from a Baha Attract to an Osia system performed audiometric and quality of life assessments with different sound processor at fitting, 3, 6 and 12 months. The speech performance in noise with the Osia system at 12 months, expressed as dB SNR, was compared to the baseline condition (Baha 5). Results The PTA4 hearing thresholds improved from 42.6 ± 11 with Baha 5 Power to 34.8 ± 13.3 with Baha 5 SuperPower and to 25.4 ± 3.5 with the Osia at 12 months, leading to a significant functional gain of 17.2 ± 10.9 dB vs Baha 5 ( p < 0.02). Speech understanding in both quiet and noise was clinically improved reaching a mean SNR of less than 1 dB at 12 months with the Osia system. Quality of life outcomes improved by more than 20% at 12-month follow-up. Conclusion Aided hearing thresholds and speech understanding in noise improved when patients were converted from the Baha Attract system to the Osia system. The aided hearing threshold was maintained up to 8 kHz (26 dB) with the Osia system.
更多
查看译文
关键词
Osia,Active transcutaneous bone conduction implant,Hearing outcomes
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要