Safety and efficacy of atrial fibrillation ablation guided by Ablation Index module
Journal of Interventional Cardiac Electrophysiology(2018)
摘要
Purpose Reconnection of pulmonary veins (PVs) remains common following radiofrequency catheter ablation for atrial fibrillation (AF). Ablation Index (AI) is a novel ablation quality marker that incorporates stability, contact force (CF), time, and power in a weighted formula. Its use seems to improve lesion durability. This is a prospective, single-arm registry to investigate on the safety and mid-term efficacy of AF ablation guided by the AI. Methods One hundred fifty-six consecutive patients (mean age 58 ± 10 years, 49% males, 44% with structural heart disease) referred for paroxysmal (124) or persistent (32) AF underwent antral PV isolation using a surround flow CF-sensing catheter guided by the AI. Radiofrequency was delivered targeting interlesion distance ≤ 6 mm and Ablation Index of 330–350 at posterior wall and 400–450 at anterior wall. Results Mean overall procedure time was 95 ± 30 min with a mean fluoroscopy time of 5 ± 6 min. Mean ablation time was 26 ± 10 min, 627/628 targeted PV were isolated. One pericardial effusion and two groin hematomas were reported; none required intervention. During a mean follow-up of 14 ± 6 months, 17 (10.8%) (9% paroxysmal AF vs 22% persistent AF, p = 0.09) patients had an atrial arrhythmia recurrence. Conclusions PV ablation guided by AI resulted feasible, achieving a high rate of isolated PVs, with a low complication rate, and allowed a high single-procedure arrhythmia-free survival at 14 months.
更多查看译文
关键词
Atrial fibrillation,Catheter ablation,Contact force,Ablation Index,Safety,Mid-term outcome
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要