Effect of Magnesium Chloride in Supporting Electrolyte for Enhancing Sensitive and Selective Electrochemical Sensor: An Approach for Anti-Rheumatic Sulfasalazine Detection

JOURNAL OF THE ELECTROCHEMICAL SOCIETY(2023)

引用 0|浏览9
暂无评分
摘要
In this work, an electroanalytical evaluation for voltammetric sensing of the anti-rheumatic sulfasalazine (SSZ) at an unmodified screen-printed graphene electrode (SPGE) is demonstrated. By using the differential pulse (DPV) technique, the SSZ produced a well-defined peak of around -0.3 V (vs Ag AgCl-1) in Britton-Robinson (BR) buffer pH 4. Supporting electrolytes, pH, and salts all significantly impact SSZ reduction. Therefore, their impact on the working solutions was assessed. We discovered that using a mixture of Britton-Robinson (BR) buffer with pH 4 and MgCl2 as a supporting electrolyte can enhance SSZ sensitivity by approximately 1.7 times while simultaneously increasing detection selectivity. Under optimal conditions, the proposed assay demonstrated the ultrasensitive determination of SSZ with a broad linear detection range from 0.01 to 100 mu M and a low detection limit of 4.7 nM (S/N = 3). To demonstrate the impact of the proposed method, the sensor has been successfully applied for the quantitative determination of SSZ in pharmaceutical, urine, and artificial serum sample. Therefore, this approach could offer simplicity, and rapidity, and serve as an alternative to the SSZ detection in practical applications. (C) 2023 The Electrochemical Society ("ECS"). Published on behalf of ECS by IOP Publishing Limited.
更多
查看译文
关键词
Sulfasalazine,Screen-Printed Graphene Electrode,Electrolyte Components,Drug,Urine,Serum
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要