Long term experience with perfluorobutane in COMPASS RICH

NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT(2023)

Cited 0|Views10
No score
Abstract
COMPASS RICH-1 has used high-purity perfluorobutane as radiator gas since 2001. The operation and control of the radiator gas has evolved over years with continuous improvements. We report on the experience gained in the 20 year-long operation of perfluorobutane as COMPASS RICH radiator.Very accurate values for the radiator gas refractive index are needed for high-performance particle identification. The procedure has evolved over years and the one presently in use, which provides refractive index estimate at the 1 ppm level, is discussed. Perfluorobutane procurement is becoming challenging, and the minimization of material waste is now a priority for the protection of the environment. Commercially available perfluorobutane needs dedicated filtering before usage and typical material losses in the filtering procedure were around 30%. Recent efforts allowed us to reduce them to about 5%. A potential alternative to fluorocarbon radiators in gaseous RICHes is also presented.
More
Translated text
Key words
COMPASS,RICH,Radiator gas,Fluorocarbon
AI Read Science
Must-Reading Tree
Example
Generate MRT to find the research sequence of this paper
Chat Paper
Summary is being generated by the instructions you defined