Reagent Effects on the Activated Partial Thromboplastin Time Clot Waveform Analysis: A Multi-Centre Study.

Wan Hui Wong,Chuen Wen Tan, Nabeelah Binti Abdul Khalid, Nadjwa Zamalek Dalimoenthe,Christina Yip,Chaicharoen Tantanate, Rodelio D Lim, Ji Hyun Kim,Heng Joo Ng

Diagnostics (Basel, Switzerland)(2023)

引用 0|浏览0
暂无评分
摘要
(1) Background: The activated partial thromboplastin time (APTT)- based clot waveform analysis (CWA) quantitatively extends information obtained from the APTT waveform through its derivatives. However, pre-analytical variables including reagent effects on the CWA parameters are poorly understood and must be standardized as a potential diagnostic assay. (2) Methods: CWA was first analysed with patient samples to understand reagent lot variation in three common APTT reagents: Pathromtin SL, Actin FS, and Actin FSL. A total of 1055 healthy volunteers were also recruited from seven institutions across the Asia-Pacific region and CWA data were collected with the Sysmex CS analysers. (3) Results: CWA parameters varied less than 10% between lots and the linear mixed model analysis showed few site-specific effects within the same reagent group. However, the CWA parameters were significantly different amongst all reagent groups and thus reagent-specific 95% reference intervals could be calculated using the nonparametric method. Post-hoc analysis showed some degree of influence by age and gender with weak correlation to the CWA (r < 0.3). (4) Conclusions: Reagent type significantly affects APTT-based CWA with minimal inter-laboratory variations with the same coagulometer series that allow for data pooling across laboratories with more evidence required for age- and gender-partitioning.
更多
查看译文
关键词
activated partial thromboplastin time,clot,coagulation,reagent,reference interval,waveform analysis
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要