Radiation dose considerations in digital radiography with an anti-scatter grid: A study using adult and pediatric phantoms

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS(2023)

引用 1|浏览1
暂无评分
摘要
BackgroundWhen using an anti-scatter grid, a decrease in receptor dose caused by its X-ray absorption seems to lead to the misperception that radiation dose needs to be increased even in digital radiography (DR). ObjectiveTo demonstrate that there is no need to increase radiation dose in DR with a grid, based on a visual evaluation using an adult and a pediatric abdomen phantom (P-AD and P-PD, respectively). Materials and methodsPhantom images with and without a grid were obtained with exposure parameters determined based on a preliminarily measured signal-to-noise ratio improvement factor (SIF), an index for potential dose reduction when using a grid. In visual evaluation, four radiologists compared phantom images with a grid applied at different dose reduction rates (0% [no reduction], 18%, 36%, and 59% for P-AD and 0% and 11% for P-PD) against an image without a grid at the baseline dose (as the reference). They graded the overall image quality of the former relative to that of the latter (reference) on a 3-point scale (3 = better, 2 = almost equal, 1 = worse). ResultsThe mean scores for dose reduction rates of 0%, 18%, 36%, and 59% were 3.00, 3.00, 2.75, and 1.00, respectively, for P-AD; those for 0% and 11% were 2.13 and 1.63, respectively, for P-PD. These results support the validity of our view that no dose increase is necessary when using an anti-scatter grid. Actually, there is even a potential for improvement in image quality with dose reduction rates of & LE;36% for P-AD. ConclusionIt is worth reconsidering the necessity of increasing radiation dose in the DR imaging of the adult and pediatric abdomens with an anti-scatter grid.
更多
查看译文
关键词
digital radiography,pediatric phantoms,radiation
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要