The accuracy of an accelerometer-based portable navigation system for total hip arthroplasty using 3D CT measurement in the supine position: a prospective multicenter study

Journal of Joint Surgery and Research(2023)

Cited 0|Views5
No score
Abstract
•This multicenter prospective study evaluated the accuracy of the navigation system.•324 THAs using supine position in 9 hospitals were prospectively enrolled.•Cup alignment was measured on postoperative 3D-CT images.•The accuracy was 3 ​± ​3°for cup inclination and 3 ​± ​3° for cup anteversion.•The patients' factor, surgical factor, or pelvic movement did not affect the accuracy.
More
Translated text
Key words
Accelerometer-based portable navigation system,Total hip arthroplasty,Supine position,Accuracy