Abstract
In this study, we aimed to examine the effect of varying β-values in the block sequential regularized expectation maximization (BSREM) algorithm under differing lesion sizes to determine an optimal penalty factor for clinical application. The National Electrical Manufacturers Association phantom and 15 prostate cancer patients were injected with 68Ga-PSMA and scanned using a GE Discovery IQ PET/CT scanner. Images were reconstructed using ordered subset expectation maximization (OSEM) and BSREM with different β-values. Then, the background variability (BV), contrast recovery, signal-to-noise ratio, and lung residual error were measured from the phantom data, and the signal-to-background ratio (SBR) and contrast from the clinical data. The increment of BV using a β-value of 100 was 120.0
More
Translated text
Key words
Image reconstruction,PET-CT scan,Prostate cancer,Bayesian approach