A dual-centre study on the radioprotective effect of a novel X-ray protection device during coronary intervention.

J Ma, K Liu, W Chen,T Wang, Z Xu, Y Li, B Zhao,L Zhou,F Wang,C Li

Clinical radiology(2023)

引用 0|浏览4
暂无评分
摘要
AIM:To investigate the shielding efficiency of a novel X-ray protection device (NPD) compared with the traditional lead clothing (TLC) during coronary intervention. MATERIALS AND METHODS:This study was performed prospectively in two centres. A total of 200 coronary interventions were included and assigned equally into the NPD or TLC group. The NPD is a floor-standing X-ray protection device, which mainly composes of a barrel-like frame and two layers of lead rubber. Thermoluminescent dosimeters (TLDs) were adopted to detect the cumulative absorbed doses, and were attached outside the NPD or TLC or body of the first operator at four different height levels in four directions during the procedure. RESULTS:The cumulative doses outside the NPD were comparable to that of the TLC (2,398.33 ± 2,341.64 versus 1,624.09 ± 1,732.20 μSv, p=0.366), and the cumulative doses inside the NPD were significantly lower than those inside the TLC (40 ± 0 versus 732.28 ± 919.83 μSv, p<0.001). As the TLC did not cover the calf segment of the operator, the area at 50 cm height from the floor in the TLC group was unshielded. The shielding efficiency of NPD was significantly higher than that of the TLC (98.2 ± 0.63% versus 52.11 ± 38.97%, p=0.021). CONCLUSION:The NPD has a significantly higher shielding efficacy than that of the TLC, in particular, it protects the operators' lower limb, liberates their lower body from wearing heavy lead apron, and may consequently reduce the radiation or body-load associated complications.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要