Effect of the surface geometry of uncemented acetabular cups on primary stability in a model of uncemented total hip replacement in dogs.
American journal of veterinary research(2023)
Abstract
Our results suggest that cups with a peripheral design (B, C) have less primary stability than hemiellipsoidal cups (A). Furthermore, models with peripheral fins (B, C) appeared to have incomplete seating if a higher force was not used during implantation and, therefore, the risk of malpositioning was increased. These data indicate that hemiellipsoidal cups provide the same or better initial stability and require a lower impaction force.
MoreTranslated text
Key words
uncemented acetabular cups,uncemented total hip replacement,primary stability