Diagnostic Accuracy of Somatosensory Evoked Potentials and Electroencephalography During Endovascular Treatment of Unruptured Cerebral Aneurysms

World Neurosurgery(2023)

引用 0|浏览3
暂无评分
摘要
OBJECTIVE:Endovascular treatment (EVT) of unruptured cerebral aneurysms (UCA) offers a safer alternative to clipping. However, it is still associated with an increased risk for Postprocedural Neurological deficit (PPND). Prompt recognition using intraoperative neurophysiologic monitoring (IONM) and intervention can reduce the incidence and impact of new postoperative neurological complications. We aim to evaluate the diagnostic accuracy of IONM in predicting PPND after EVT of UCA. METHODS:We included 414 patients who underwent EVT for UCA from 2014 to 2019. The sensitivities, specificities, and diagnostic odds ratio of somatosensory evoked potentials and electroencephalography monitoring methods were calculated. We also determined their diagnostic accuracy using receiver operating characteristic plots. RESULTS:The highest sensitivity of 67.7% (95% confidence interval {CI}, 34.9%-90.1%) was obtained when either modality had a change. Simultaneous changes in both modalities have the highest specificity of 97.8% (95% CI, 95.8%-99.0%). The area under the receiver operating characteristic curve was 0.795 (95% CI, 0.655-0.935) for changes in either modality. CONCLUSIONS:IONM with somatosensory evoked potentials alone or in combination with electroencephalography has high diagnostic accuracy in detecting periprocedural complications and resultant PPND during EVT of UCA.