Accuracy and Reproducibility of Assessing Patent Foramen Ovale on Coronary Computed Tomographic Angiography: A Comparison with Transesophageal Echocardiography

Research Square (Research Square)(2021)

引用 0|浏览2
暂无评分
摘要
Abstract This study was undertaken to determine if coronary computed tomographic angiography (CCTA) can help to assess patent foramen ovale (PFO) with high accuracy and reproducibility when compared to Transesophageal Echocardiography (TEE). In total, 75 patients (31 men, 44 women; mean age, 45 ± 9 years) with suspected PFO were evaluated using coronary CCTA and TEE. PFO tunnel length (TL) and the opening diameter of the left atrial entrance (ODLAE) and right atrial entrance (ODRAE), as well as contrast shunt (if present due to PFO), were measured by both modalities. PFO was detected in 67 patients with TEE. The sensitivity for the detection of PFO with CCTA was 85.3%; specificity, 71.4%; positive predictive value, 96.7%; and negative predictive value, 33.3%. Both modalities demonstrated good agreement in measuring TL and ODLAE of PFO. However, the ODRAE of TEE was different from that of CCTA (1.14 ± 0.4 mm and 1.45 ± 0.5 mm, respectively, p = 0.04). The intraobserver and interobserver variability and agreement for TL, ODRAE, and ODLAE of PFO were excellent between the two measurements. CCTA provided a method for detection of PFO with high accuracy and reproducibility compared with TEE. Therefore, CCTA is a practical and efficient alternative to TEE for PFO diagnosis.
更多
查看译文
关键词
assessing patent foramen ovale,coronary computed tomographic angiography,echocardiography