Unusual atrial pacing induced by type II far-field p-wave sensing in a patient with complete atrioventricular block

PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY(2024)

引用 0|浏览2
暂无评分
摘要
An 88-year-old Japanese woman underwent DDD pacemaker (MicroPort KORA 250 DR, V lead: VEGA R52) implantation for complete atrioventricular block. A 12-lead electrocardiogram for a routine examination showed atrial pacing within the intrinsic P wave, followed by inhibition of ventricular pacing. Pacemaker interrogation revealed no abnormalities in the basic parameters; however, ventricular pacing was inhibited by far-field sensing of intrinsic atrial waves before atrial events; type II far-field P-wave sensing. As a result, unusual atrial pacing occurred due to the pause suppression algorithm, which is the one of the functions that prevent atrial fibrillation development.
更多
查看译文
关键词
complete atrioventricular block,crosstalk,far-field P-wave (FFPW) sensing,pause suppression,type II far-field