Screening Clones for Monoclonal Antibody Production Using Droplet Microfluidics Interfaced to Electrospray Ionization Mass Spectrometry.

Journal of the American Society for Mass Spectrometry(2023)

引用 0|浏览0
暂无评分
摘要
As one of the most critical steps in process development for protein therapeutics, clone selection and cell culture optimization require a large number of samples to be screened for high titer and desirable molecular profiles. Typical analytical techniques, such as chromatographic approaches, often take minutes per sample which are inefficient for large-scale screenings. Droplet microfluidics coupled to mass spectrometry (MS) represents an attractive approach due to its low volume requirements, high-throughput capabilities, label-free nature, and ability to handle complex mixtures. In this work, we coupled a modified protein cleanup protocol with a droplet-MS workflow for mAb titer screening to guide clone selection. With this droplet approach we achieved a throughput of 0.04 samples/s with an LoD of 0.15 mg/mL and an LoQ of 0.45 mg/mL. To test its performance in a real-world setting, this workflow was applied to a 35-clone screen, where the top 20% producing clones were identified. In addition, we coupled our sample cleanup protocol to a high-resolution MS and compared the glycan profiles of the high titer clones. This work demonstrates that droplet-MS provides a rapid way of clone screening and cell culture optimization based on titer and molecular structure of the expressed proteins. Future work is aimed at increasing the throughput and automation of this droplet-MS technique.