Development of an optimised method for measurement of iodine-129 in decommissioning wastes using ICP-MS/MS

JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY(2023)

Cited 0|Views13
No score
Abstract
Accurate, low-level measurement of the long-lived fission product I-129 is important for waste characterisation and long-term monitoring of waste facilities and the surrounding environment. Inductively coupled plasma mass spectrometry (ICP-MS) is well-suited to high throughput, sensitive measurement of I-129 as a cost-effective alternative to other mass spectrometric and decay counting techniques. Accurate I-129 measurement by ICP-MS is affected by the multiple interferences on m/z = 129, necessitating multi-stage sample preparation and/or mathematical correction. This study assesses the capabilities of tandem ICP-MS/MS for rapid and routine measurement of I-129 in nuclear wastes. The advantages of the tandem setup for removal of isobaric, polyatomic and tailing interferences are demonstrated, as are the improvements in sensitivity through matrix modification and the importance of selecting an appropriate internal standard. The optimised setup was applied to measurement of various decommissioning waste simulants following direct extraction of I-129. The procedure achieved an instrument detection limit of 1.05 x 10(-4) Bq g(-1) (0.017 ng g(-1)) for I-129, which is two orders of magnitude below the target out-of-scope limit of 0.01 Bq g(-1) (1.57 ng g(-1)), with good agreement between ICP-MS/MS and liquid scintillation counting (LSC). The results show that rapid, routine, low-level measurement of I-129 is achievable by ICP-MS/MS for end users in decommissioning and environmental monitoring.
More
Translated text
AI Read Science
Must-Reading Tree
Example
Generate MRT to find the research sequence of this paper
Chat Paper
Summary is being generated by the instructions you defined