Frequency Response of a Six-Electrode MET Sensor at Extremely Low Temperatures

Sensors(2023)

Cited 0|Views4
No score
Abstract
Four-electrode electrochemical cells are widely used for signal conversion in molecular-electronic transfer (MET) motion sensors. The most used ACCA (anode-cathode-cathode-anode) configuration has proven its performance and usefulness for obtaining a superior conversion factor and a wider frequency range over standard geophones at room temperature. However, the MET sensor conversion factor decreases a thousand-fold or more when the temperature drops from room temperature to 233 K. In the design suggested is this paper, a pair of additional gate (G) electrodes has been added outside the standard ACCA cell. An experimental study of the temperature behavior of the resulting G-ACCA-G six-electrode configuration showed that the effects of temperature changes on the cell conversion factor are 5.2 times weaker compared with the standard ACCA configuration.
More
Translated text
Key words
low temperatures,sensor,frequency,six-electrode