Safety, Feasibility of Controllable Decrease of Vena Cava Pressure by Doraya Catheter in Heart Failure

JACC-BASIC TO TRANSLATIONAL SCIENCE(2023)

引用 0|浏览15
暂无评分
摘要
Lowering elevated central venous pressure may reduce renal dysfunction in acute heart failure (AHF) patients. The Doraya catheter lowers renal venous pressure by creating a gradient in the inferior vena cava below the renal veins. Here, we present a first-in-human feasibility study of the Doraya catheter performed on 9 AHF patients. We assessed the safety, feasibility, and acute clinical (hemodynamic and renal) effects of transient Doraya catheter deployment when added to the standard diuretic-based regimen in AHF patients with a poor diuretic response. The procedures decreased central venous pressure from 18.4 +/- 3.8 mm Hg to 12.4 +/- 4.7 mm Hg (P < 0.001) and improved mean diuresis and clinical signs of congestion. No device-related serious adverse events were observed. Thus, Doraya catheter deployment was safe and feasible in AHF patients. (First In Human Study of the Doraya Catheter for the Treatment of AHF Patients; NCT03234647)(c) 2023 The Authors. Published by Elsevier on behalf of the American College of Cardiology Foundation. This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
更多
查看译文
关键词
vena cava pressure,doraya catheter,heart failure
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要