Noninvasive measurement of temperature for simulated tissue based on piezoelectric micromachined ultrasonic transducers

JOURNAL OF MICROMECHANICS AND MICROENGINEERING(2023)

引用 0|浏览11
暂无评分
摘要
This paper reports a noninvasive measurement of temperature for simulated tissue based on an AlN piezoelectric micromachined ultrasonic transducer array. Polydimethylsiloxane and a built-in heating plate are used to simulate the thermal therapy of human tissue. Based on time-of-flight (ToF) from the simulated tissue surface to the heating plate surface and corresponding temperature data, the piecewise linear fitting method is used to fit the ToF-temperature curve, and its sensitivity is 41.5 ns degrees C-1 (below 50 degrees C) and 30.5 ns degrees C-1 (from 50 degrees C to 65 degrees C). The average error of real-time temperature measurements is 0.58 degrees C for temperatures below 65 degrees C. The proposed method can provide a good solution for temperature detection during thermal therapy.
更多
查看译文
关键词
noninvasive measurement,temperature,simulated tissue
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要