Inter‐device concordance of Veggie Meter, a reflection spectroscopy to measure skin carotenoids

Journal of Biophotonics(2023)

引用 0|浏览12
暂无评分
摘要
The Veggie Meter® (VM) is a refraction spectrometer that detects skin carotenoids. We tested the variability of the two modes (single-scan and averaging modes) of four VMs of three different versions (VM-1, 2, 3, 4) with 92 healthy volunteers. Both the modes exhibited a high intraclass correlation coefficient; however, the averaging mode had a significantly lower coefficient of variance than the single-scan mode. The Bland-Altman analysis showed a systematic error between VM-1 and other three VMs. The actual errors between the VM-1 and the other three VMs in the averaging mode were 7.4, 10.4, and 11.8% of the median of VM scores, but the errors could be reduced to 2.8, 6.3, and 7.0% by compensating VM scores using the regression equations. Averaging mode had higher accuracy than single-scan mode. The reliability of the VMs was validated by the small CV and high ICC. The error was improved using linear regression compensation.
更多
查看译文
关键词
Reflection spectroscopy,coefficient of variation,inter-device error,intra-class correlation coefficient,linear regression,skin carotenoids