[Study of Phantom Length Correction for kV-Cone Beam CT Dose Evaluation Method Using Farmer-type Ionization Chamber and Cylindrical Phantom].

Nihon Hoshasen Gijutsu Gakkai zasshi(2023)

引用 0|浏览2
暂无评分
摘要
In this study, we investigated the necessary phantom length for dose evaluation of kV-CBCT mounted on the linear accelerator using a Farmer-type ionization chamber and cylindrical PMMA phantom, which many radiotherapy facilities own. Furthermore, a phantom length correction factor was proposed to compensate for the lack of scattered radiation contributed from the craniocaudal direction due to the inadequate phantom length. The air-absorbed dose at the center of a cylindrical PMMA phantom measured by a Farmer-type ionization chamber indicated that the contribution of scattered radiation saturates when the phantom length is 300 mm or longer. The phantom length correction factor was calculated from an approximate curve showing the relationship of the charge ratio measured using phantoms of 300 mm and 150 mm in length to the X-ray beam width. The air-absorbed dose measured with the 150-mm length phantom, corrected by the phantom length correction factor, showed a 1.61% dose difference from the air-absorbed dose measured with the 300-mm length phantom. In this study, the air-absorbed dose at the center of the phantom could be estimated over a wide X-ray beam width only using a 150-mm length cylindrical PMMA phantom. The method proposed in this study could be used in any radiation therapy facility.
更多
查看译文
关键词
Farmer-type ionization chamber,cone-beam computed tomography (CBCT),dose evaluation,image-guided radiotherapy (IGRT),phantom length
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要