Initial safety outcomes for the Rigicon Infla10 (R) inflatable penile prosthesis

BJU INTERNATIONAL(2023)

引用 3|浏览11
暂无评分
摘要
ObjectiveTo analyse data from patient information forms (PIFs) submitted to the manufacturer of a new three-piece inflatable penile prosthesis (IPP), the Rigicon Infla10 (R), to summarize early outcomes. MethodsA retrospective review of PIFs from 319 patients implanted with the IPP between 6 January 2019 and 31 December 2021 was performed to assess device durability and rates of reoperation for infection, mechanical failure or medical reasons. ResultsThe mean +/- sd (range) follow-up was 21.2 +/- 11 (5-36) months and the mean +/- sd patient age was 58.5 +/- 8.7 years. Of the total, 4.4% of devices required removal or revision. Complications included mechanical failure (2.5%) and component migration (0.7%). No infections were observed in this series. One patient (0.4%) required revision for inadequate straightening of Peyronie's disease. Three patients (0.9%) requested device removal because of dissatisfaction. A total of 95.6% of the Rigicon Infla10 devices were free from explant or revision 21 months after the original implant date. Kaplan-Meier analysis showed the rates of cumulative survival of the device at 12, 24, 36 months were 95.6%, 94.7% and 93.7%, respectively. ConclusionsThe vast majority of the early Rigicon Infla10 IPPs implanted prior to January 2022 were included in this retrospective analysis of volunteered PIFs. These early results demonstrate initial durability from reoperation equivalent to that of other contemporary devices.
更多
查看译文
关键词
inflatable penile prosthesis,safety,Rigicon Infla10 (R),three-piece inflatable,penile implant
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要