Partly-O-Diamond Solution-Gate Field-Effect Transistor as an Efficient Biosensor of Glucose

JOURNAL OF THE ELECTROCHEMICAL SOCIETY(2023)

Cited 1|Views13
No score
Abstract
Assessment of glucose concentration is vital for diagnosis and treatment of diabetes mellitus. Since the electrochemical advantages of diamond for biosensing, we introduced partly-oxygen-diamond solution-gate field-effect transistor (SGFET) to detect glucose. Partly-O-diamond was obtained from H-diamond by ultraviolet ozone treatment. Then, 1-pyrenebutyric acid-N-hydroxy succinimide ester and glucose oxidase were modified on diamond surface. At last, shifts of transfer characteristics were applied to determine glucose concentration. The as-prepared biosensor exhibited a wide linear response from 10(-5) to 10(-1 )M with high sensitivity of -53 mV/log(10) [glucose concentration]. Furthermore, this biosensor showed good repeatability and stability in multiple detections.