Development of digital electronics for the search of SHE nuclei using GARIS-II/III at RIKEN

Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment(2023)

引用 0|浏览14
暂无评分
摘要
This work describes the implementation of a digital data acquisition system for the search and study of short-lived isotopes or superheavy elements at the GARIS-III separator at the RIKEN Nishina Center. A reduction to 26 +/- 0.5 keV in the average value of the Full Width at Half Maximum of the alpha spectrum at 7.133 MeV was achieved, combined with a reduction of the dead time of the overall system to around a few tens of nanoseconds for the detection of very fast decay chains. Moreover, we measured a reduction of the background induced by particles crossing the implantation detector using pulse shape analysis techniques. The detail of the optimization and analysis will be presented.
更多
查看译文
关键词
Super heavy elements,Digital electronics,Pulse shape analysis,Silicon detector