Successful uniportal thoracoscopic removal of a new generation implantable loop recorder accidentally migrated into the left pleural cavity and concomitant re-implantation: A case report.

International journal of surgery case reports(2023)

引用 1|浏览0
暂无评分
摘要
In case of intrathoracic migration of ILRs, early removal by mini-invasive approach is recommended as well as concomitant re-implantation. Beyond periodic monitoring of ILRs by cardiologists, strict radiological follow-up with chest X-ray is advisable after implantation, in order to early identify any abnormalities and correctly manage them.
更多
查看译文
关键词
Case report,Implantable loop recorder (ILR),Intrathoracic migration,Re-implantation,Surgical removal,Uniportal video-assisted thoracic surgery