Simultaneous voltammetric sensing of Zn 2+ , Cd 2+ , and Pb 2+ using an electrodeposited Bi-Sb nanocomposite modified carbon paste electrode.

RSC advances(2023)

引用 0|浏览3
暂无评分
摘要
A sensor for detecting Zn, Cd, and Pb ions simultaneously based on the square wave anodic stripping response at a bismuth antimony (Bi-Sb) nanocomposite electrode was developed. The electrode was prepared by electrodepositing bismuth and antimony on the surface of a carbon-paste electrode (CPE) while also reducing the analyte metal ions. The structure and performance of the Bi-Sb/CPE electrode were studied using scanning electron microscopy, X-ray diffraction, electrochemical impedance spectroscopy, and cyclic voltammetry. Operational conditions including the concentration of Sb and Bi, the type of electrolyte, pH, and preconcentration conditions were optimized. The linear ranges were determined to be 5-200 μg L for Zn, 1-200 μg L for Cd, and 1-150 μg L for Pb with the optimized parameters. The limits of detection were 1.46 μg L, 0.27 μg L, and 0.29 μg L for Zn, Cd, and Pb, respectively. Furthermore, the Bi-Sb/CPE sensor is capable of selective determination of the target metals in the presence of the common cationic and anionic interfering species (Na, K, Ca, Mg, Fe, Mn, Co, Cl, SO and HCO ). Finally, the sensor was successfully applied to the simultaneous determination of Zn, Cd, and Pb in a variety of real-world water samples.