Imprecision remains to be improved in the measurement of serum cystatin C with heterogeneous systems

Clinical chemistry and laboratory medicine(2023)

引用 0|浏览12
暂无评分
摘要
Objectives: Except for the large bias of some measurement systems for serum cystatin C (CysC) measurements, unacceptable imprecision has been observed for the heterogenous system. This study analyzed the external quality assessment (EQA) results in 2018-2021 to provide an insight into the imprecision of CysC assays.Methods: Five EQA samples were sent to participating laboratories every year. Participants were divided into reagent/calibrator-based peer groups, for which the robust mean of each sample and robust coefficient of variation (CV) were calculated by Algorithm A from ISO 13528. Peers with more than 12 participants per year were selected for further analysis. The limit of CV was determined to be 4.85% based on clinical application requirements. The concentration-related effect on CVs was investigated using logarithmic curve fitting; the difference in medians and robust CVs between instrument-based subgroups was also evaluated.Results: The total number of participating laboratories increased from 845 to 1,695 in four years and heterogeneous systems remained the mainstream (>= 85%). Of 18 peers with >= 12 participants, those using homogeneous systems showed relatively steady and small CVs over four years, with the mean four-year CVs ranging from 3.21 to 3.68%. Some peers using heterogenous systems showed reduced CVs over four years, while 7/15 still had unacceptable CVs in 2021 (5.01-8.34%). Six peers showed larger CVs at the low or high concentrations, and some instrument-based subgroups presented greater imprecision than others. Conclusions: More efforts should be made to improve the imprecision of heterogeneous systems for CysC measurement.
更多
查看译文
关键词
cystatin C,external quality assessment,imprecision,peer group evaluation
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要