Cross-Calibration of iDXA and pQCT Scanners at Rural and Urban Research Sites in The Gambia, West Africa

Calcified tissue international(2023)

Cited 0|Views21
No score
Abstract
Between-scanner differences in measures of bone and body composition can obscure or exaggerate physiological differences in multi-site studies or the magnitude of changes in longitudinal studies. We conducted a cross-calibration study at two bone imaging centres in The Gambia, West Africa where DXA (dual-energy X-ray absorptiometry) and pQCT (peripheral Quantitative-Computed Tomography) are routinely used. Repeat scans were obtained from 64 Gambian adults (58% Male) aged Mean(SD) 30.9 (13.5) years with Mean(SD) body mass index (BMI) 21.7 (4.0) kg/m 2 , using DXA (GE Lunar iDXA, whole body [WB], total hip [TH], lumbar spine [LS]) and pQCT (Stratec XCT2000L/XCT2000, tibia 4%, 50% sites). Between-scanner differences were tested using paired t tests ( p < 0.05). Between-scanner correlation was explored with linear regression, and cross-calibration equations derived. Bland–Altman analysis investigated machine trend/bias. When differences were detected ( p < 0.05), cross-calibration equations were applied to urban values, with t tests and Bland Altman analysis repeated. Between-scanner differences exceeded the predefined level of statistical significance ( p < 0.05) for WB aBMD and BA; all pQCT measures vBMD, BMC, cortical cross-sectional area (CSA) and stress–strain index (SSI). Between-scanner correlation was high ( R 2 :0.92–0.99), except pQCT Mu.Den ( R 2 = 0.51). Bland Altman plots indicated bias increased with increasing BMD. Cross-calibration equations attenuated all between-scanner differences and systematic bias. Cross-calibration, particularly of pQCT scanners, is an important consideration in multi-site studies particularly where between population comparisons are intended. Our experiences and findings may be generalisable to other resource-limited settings where the logistics of sourcing parts and in-country repair may result in lengthy scanner downtime.
More
Translated text
Key words
Body composition,Cross-calibration,DXA,Densitometry,pQCT
AI Read Science
Must-Reading Tree
Example
Generate MRT to find the research sequence of this paper
Chat Paper
Summary is being generated by the instructions you defined