Comparison of 50° handheld fundus camera versus ultra-widefield table-top fundus camera for diabetic retinopathy detection and grading

Eye (London, England)(2023)

引用 0|浏览10
暂无评分
摘要
Objectives To compare the performance of a handheld fundus camera with standard 50° visual field to ultra-widefield (UWF) table-top fundus camera in diabetic retinopathy (DR) detection and grading. Methods Patients affected by diabetes mellitus and referred to our diabetic retinopathy clinic were enroled and underwent fundus photography in mydriasis. All photos were taken using the ultra-widefield table-top fundus camera Zeiss Clarus™ 500 (four fields per eye) and the Optomed Aurora® handheld fundus camera (3 fields per eye). The following parameters were analysed: the gradability of the images, the grade of DR, and diabetic maculopathy (DM), the presence of hypertensive retinopathy (HR) and the presence of other ocular diseases. Results We enroled 759 eyes of 384 diabetic patients and analysed 5313 fundus photos. The handheld fundus camera obtained a sensitivity of 84.2% and specificity of 95.4% for referable cases. Moreover, it obtained, compared to UWF, an almost perfect agreement with linear weighting for DR, DM and HR ( k = 0.877, k = 0.854, and k = 0.961, respectively). The lowest sensitivity was achieved for proliferative DR (58.7% sensitivity, 100% specificity). Conclusions Optomed Aurora® handheld fundus camera imaging showed a strong agreement compared to UWF in grading DR, considering all DR and DM grades, in mydriasis. However, the use of UWF imaging increases the detection of referable eyes.
更多
查看译文
关键词
diabetic retinopathy detection,fundus camera,ultra-widefield,table-top
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要