Experimental investigation of improved tolerance for concentration mismatch in potassium isotope analysis on a hexapole collision cell MC-ICP-MS (Nu Sapphire)

JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY(2023)

引用 1|浏览10
暂无评分
摘要
Collision cell multi-collector inductively coupled plasma mass spectrometry (CC-MC-ICP-MS) has made routine high precision potassium (K) isotope analysis possible using sample-standard bracketing. However, as reported in previous studies, strict concentration match between a sample and bracketing standard (<2% tolerance reported from most labs worldwide) is often required to avoid analytical bias. This small tolerance for concentration mismatch would naturally lead to low analytical efficiency. To better characterize and potentially improve this, we conducted a series of experiments on the hexapole parameters of the Nu Sapphire CC-MC-ICP-MS. The results reveal that RF Ref. (the voltage of the RF alternating current applied to the hexapole rods) is the primary parameter that controls the concentration mismatch effect. For a set RF Bias Ref. value (the potential for the hexapole) of -0.8 V, the optimal RF Ref. value of 1.05 V identified by the flattest part in the K-41/K-39 ratio versus RF Ref. plot has increased the tolerance for K concentration mismatch to 30% from 4% at a RF Ref. of 1.5 V. The cell gas flow rate is another important parameter and optimal values of 2 sccm and 5 sccm are recommended for He and H-2, respectively. A much higher tolerance for concentration mismatch using optimized parameters has greatly improved the efficiency of high precision K isotope analyses using CC-MC-ICP-MS and will open the possibility for in situ K isotopic measurements.
更多
查看译文
关键词
potassium isotope analysis,concentration mismatch,nu sapphire,mc-icp-ms
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要