Functionalization of the Implant Surface Made of NiTi Shape Memory Alloy.

Materials (Basel, Switzerland)(2023)

引用 1|浏览4
暂无评分
摘要
To functionalize and improve the biocompatibility of the surface of a medical implant made of NiTi shape memory alloy and used in practice, a clamp, multifunctional layers composed of amorphous TiO interlayer, and a hydroxyapatite coating were produced. Electrophoresis, as an efficient method of surface modification, resulted in the formation of a uniform coating under a voltage of 60 V and deposition time of 30 s over the entire volume of the implant. The applied heat treatment (800 °C/2 h) let toa dense, crack-free, well-adhered HAp coating with a thickness of ca. 1.5 μm. and a high crack resistance to deformation associated with the induction of the shape memory effect in the in the deformation range similar to the real implant work after implantation. Moreover, the obtained coating featured a hydrophilic (CA = 59.4 ± 0.3°) and high biocompatibility.
更多
查看译文
关键词
NiTi shape memory implant,electrophoretic deposition (EPD),hydroxyapatite (HAp),surface functionalization