Clinical commissioning of the first point-of-care spectral photon-counting CT for the upper extremities.

Medical physics(2023)

引用 0|浏览7
暂无评分
摘要
A routine QA framework for SPCCT systems has been proposed. Image quality and radiation dose were assessed using newly designed phantoms, relevant metrics, and automated algorithms. Baseline values were established and tolerance levels discussed for the MARS SPCCT scanner based on collected data and international recommendations.
更多
查看译文
关键词
MARS Extremity 5X120,commissioning,image quality,quality assurance,radiation dose,spectral photon-counting CT