Influence of surgical position and registration methods on clinical accuracy of navigation systems in brain tumor surgery.

Scientific reports(2023)

Cited 3|Views20
No score
Abstract
The aim of this study was to evaluate the influence of skin distortion due to surgical positioning on the clinical accuracy of the navigation system. The distance errors were measured in four fiducial markers (anterior, posterior, right, and left of the head) after the registration of the navigation system. The distance errors were compared between the surface-merge registration (SMR) method using preoperative imaging and the automatic intraoperative registration (AIR) method using intraoperative imaging. The comparison of the distance errors were performed in various surgical positions. The AIR method had the significant accuracy in the lateral markers than the SMR method (lateral position, 3.8 mm vs. 8.95 mm; p < 0.0001; prone position, 4.5 mm vs. 13.9 mm; p = 0.0001; 5.2 mm vs. 11.5 mm; p = 0.0070). The smallest distance errors were obtained close to the surgical field in the AIR method (3.25-3.85 mm) and in the forehead in the SMR method (3.3-8.1 mm). The AIR method was accurate and recommended for all the surgical positions if intraoperative imaging was available. The SMR method was only recommended for the supine position, because skin distortion was frequently observed in the lateral region.
More
Translated text
Key words
brain tumor surgery,surgical position,navigation systems,clinical accuracy