Spatial resolution, noise properties, and detectability index of a deep learning reconstruction algorithm for dual-energy CT of the abdomen.

Medical physics(2023)

引用 4|浏览1
暂无评分
摘要
DE imaging with DLR maintained spatial resolution and reduced noise magnitude while displaying less change in noise texture than IR. The d' was also higher with DLR than IR, suggesting superiority in detectability of iodinated contrast. Despite these trends being consistent with those previously established for SE imaging, there were some noteworthy differences. For DE imaging there was no improvement in resolution compared to IR and a change in noise texture. DE imaging with low keV and DLR had superior detectability to SE DLR at the high dose but was not better than SE-80 kV at low dose.
更多
查看译文
关键词
DECT,deep learning image reconstruction,image quality