Predicting children's real-ear-to-coupler differences based on tympanometric data

International journal of audiology(2023)

引用 1|浏览0
暂无评分
摘要
ObjectivePaediatric hearing-aid verification relies on measures of output obtained from the ear canal or in a coupler with the child's real-ear-to-coupler difference (RECD). Measured RECD cannot always be completed in children, leading to fitting inaccuracies. Audiologists often have tympanometry data that characterises the child's ear-canal acoustics. The goal of this study was to determine if tympanometry can be used to improve predictions of measured RECD.DesignA retrospective analysis of RECD and admittance, tympanometric peak pressure, and equivalent ear-canal volume from 226 Hz tympanometry collected as part of a longitudinal study of children with hearing loss were modelled with Bayesian hierarchical regression.Study sampleTwo-hundred sixty-six children with mild-to-severe hearing loss contributed data.ResultsAge-based average RECD models were within 3 dB of measured RECD values in 54% of cases with normal middle ear status and 50.6% of cases with abnormal middle ear status. Immittance-predicted RECD were within 3 dB in 69.6% of cases with normal middle ear status and 74.4% of cases with abnormal middle ear status.ConclusionImmittance-predicted RECD was more accurate than age-based average RECD, particularly in children with abnormal middle ear status. The findings suggest that 226 Hz tympanometry could be used clinically to improve predictions of measured RECD when it cannot be measured.
更多
查看译文
关键词
Paediatric,hearing-aid,acoustics,real-ear-to-coupler difference,verification,hearing loss
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要