Identification of a frit-related sample carryover in newborn screening by tandem mass spectrometry

Journal of Mass Spectrometry and Advances in the Clinical Lab(2023)

Cited 0|Views15
No score
Abstract
•Sample carryover in newborn screening test can increase false positive results.•The risk of carryover is increased with newer generations of mass spectrometers.•Column-related carryover due to a stainless-steel frit inline filter was detected.•Compound-dependent factor of stainless-steel frit-related carryover is suggested.•Routine backpressure monitoring and timely replacement of frit rings are mandatory.