Mechanical extraction of implantable cardioverter-defibrillator leads with a dwell time of more than 10 years: insights from a single high-volume centre.
Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology(2023)
摘要
The TLE procedures involving the extraction of old ICD leads were effective and safe. The outcomes of ICD lead removal with a dwell time of >10 years did not differ significantly compared with younger ICD leads. However, extraction of older ICD leads required more frequent necessity for utilizing multiple extraction tools, more experience and versatility of the operator, and increased surgery costs.
更多查看译文
关键词
cardiac implantable electronic devices,complications,effectiveness,implantable cardioverter-defibrillator,lead,safety,transvenous lead extraction
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要